- MakaleYalnız künye2011Determination of proper peaking time for Ultra-LEGe detector
Karabidak, S. M.; Kaya, S.; Çevik, U.; Çelik, A.
Reducing count losses and pile-up pulse effects in quantitative and qualitative analysis is necessary for accuracy of analysis. Therefore, the proper peaking time for particular detector systems is important. The characteristic X-rays emitted from pure some elements were detected by using an Ultra-LEGe detector connecting a Tennelec TC 244 spectroscopy amplifier at different peaking time modes. Overall pulse widths were determined by an HM 203-7 oscilloscope connecting amplifier. The proper peaking time for Ultra-LEGe is determined as 3.84 mu s. (C) 2011 Elsevier Ltd. All rights reserved.