Comparison of the effects of operating microscopes with light emitting diode and halogen light source on the eye: a rabbit study

MakaleYalnız künye
  • SDG 3 · Sağlık ve Kaliteli Yaşam

Özet

Purpose To evaluate the potential toxicity of operation microscopes with halogen and light emitting diode (LED) light source on the rabbit eyes. Materials and methods Thirty-two eyes of 16 male New Zealand pigmented rabbits were involved in the study. The rabbits were divided into two groups according to the type of light source applied. Only one eye of each rabbit was exposed to illumination light, unexposed fellow eyes served as the control group. Experimental groups included group 1 exposed to halogen light for 2 h and evaluated 1 day and 1 week after the illumination, group 2 exposed to LED light for two hours and evaluated 1 day and 1 week after the illumination. On the first and seventh days after exposing the light, we evaluated the rabbit corneas using in vivo confocal microscopy (IVCM). At the end of the seventh day, the Hematoxylin-eosin staining and TUNEL staining were performed to investigate the presence of apoptosis in the retina and retina pigment epithelium. Results Early IVCM findings revealed corneal epithelial cell ovalization and indistinct intercellular borders in the halogen light group. We also observed more increase in the keratocyte density index (23.7% vs 14.1%, p = 0.001, respectively) and the Bowman reflectivity index (12.4% vs 4.1%, p = 0.001, respectively) at first day of the light exposure in halogen light group compared to LED light group. However, late IVCM indicated that these findings disappeared one week later. No apoptosis was observed in the corneal and retinal layers in early and late examination groups. Conclusion The present experimental study demonstrated that both halogen and LED lights, which were commonly used for microscopic eye surgery, had no sustained adverse effect on the cornea and retina of the rabbits; however, halogen light had a temporary adverse effect on corneal epithelium and stroma, which resolved within 1 week.

Anahtar Kelimeler

RDL Kalıcı Bağlantı (PID)
https://rdlopen.rawdatalibrary.net/r/gumushane/16403Resmi atıflarda kullanılan kalıcı adres
Tam Kayıt
AlanDeğerDil
dc.contributor.authorAydin, Bahir
dc.contributor.authorOzgur, Armagan
dc.contributor.authorOzdemir, Huseyin Baran
dc.contributor.authorInan, Mehmet Arda
dc.contributor.authorAtalay, Hatice Tuba
dc.contributor.authorBayrakceken, Kemal
dc.contributor.authorOzmen, Mehmet Cuneyt
dc.contributor.authorUcgul, Ahmet Yucel
dc.contributor.authorGocun, Pinar Uyar
dc.date.accessioned2026-10-22T10:29:18.704520482Z
dc.date.available2026-10-22T10:29:18.704520482Z
dc.date.issued2021-10
dc.description.abstractPurpose To evaluate the potential toxicity of operation microscopes with halogen and light emitting diode (LED) light source on the rabbit eyes. Materials and methods Thirty-two eyes of 16 male New Zealand pigmented rabbits were involved in the study. The rabbits were divided into two groups according to the type of light source applied. Only one eye of each rabbit was exposed to illumination light, unexposed fellow eyes served as the control group. Experimental groups included group 1 exposed to halogen light for 2 h and evaluated 1 day and 1 week after the illumination, group 2 exposed to LED light for two hours and evaluated 1 day and 1 week after the illumination. On the first and seventh days after exposing the light, we evaluated the rabbit corneas using in vivo confocal microscopy (IVCM). At the end of the seventh day, the Hematoxylin-eosin staining and TUNEL staining were performed to investigate the presence of apoptosis in the retina and retina pigment epithelium. Results Early IVCM findings revealed corneal epithelial cell ovalization and indistinct intercellular borders in the halogen light group. We also observed more increase in the keratocyte density index (23.7% vs 14.1%, p = 0.001, respectively) and the Bowman reflectivity index (12.4% vs 4.1%, p = 0.001, respectively) at first day of the light exposure in halogen light group compared to LED light group. However, late IVCM indicated that these findings disappeared one week later. No apoptosis was observed in the corneal and retinal layers in early and late examination groups. Conclusion The present experimental study demonstrated that both halogen and LED lights, which were commonly used for microscopic eye surgery, had no sustained adverse effect on the cornea and retina of the rabbits; however, halogen light had a temporary adverse effect on corneal epithelium and stroma, which resolved within 1 week.
dc.description.sponsorshipProjects of Scientific Investigation, Gazi University [01/2015-16]
dc.identifier.doi10.1080/15569527.2021.1949337
dc.identifier.eissn1556-9535
dc.identifier.endpage325
dc.identifier.issn1556-9527
dc.identifier.issue4
dc.identifier.pmid34225552
dc.identifier.startpage319
dc.identifier.urihttps://rdlopen.rawdatalibrary.net/r/gumushane/16403
dc.identifier.volume40
dc.identifier.wosWOS:000670116300001
dc.language.isoen
dc.publisherTAYLOR & FRANCIS LTD
dc.relation.ispartofCUTANEOUS AND OCULAR TOXICOLOGY
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.subjectConfocal microscopy
dc.subjecthalogen light
dc.subjectlight emitting diode
dc.subjectoperating microscopes
dc.subjecttoxicity
dc.subject.classificationOphthalmology
dc.subject.classificationToxicology
dc.subject.otherINDUCED RETINAL PHOTOTOXICITY
dc.subject.otherPHOTOCHEMICAL DAMAGE
dc.subject.otherBODY-TEMPERATURE
dc.subject.otherCATARACT-SURGERY
dc.subject.otherMACULOPATHY
dc.subject.otherRISK
dc.titleComparison of the effects of operating microscopes with light emitting diode and halogen light source on the eye: a rabbit study
dc.typeArticle
dspace.entity.typePublication
rdl.citation.openalex5
rdl.citation.wos2
rdl.oa.statusclosed
rdl.sdgSDG03
rdl.source.indexWOS
rdl.wos.indexScience Citation Index Expanded (SCI-EXPANDED)