The determination of the pulse pile-up reject (PUR) counting for X and gamma ray spectrometry

BildiriYalnız künye

Özet

The collection the charged particles produced by the incident radiation on a detector requires a time interval. If this time interval is not sufficiently short compared with the peaking time of the amplifier, a loss in the recovered signal amplitude occurs. Another major constraint on the throughput of modern x or gamma-ray spectrometers is the time required for the subsequent the pulse processing by the electronics. Two above-mentioned limitations are cause of counting losses resulting from the dead time and the pile-up. The pulse pile-up is a common problem in x and gamma ray radiation detection systems. The pulses pile-up in spectroscopic analysis can cause significant errors. Therefore, inhibition of these pulses is a vital step. A way to reduce errors due to the pulse pile-up is a pile-up inspection circuitry (PUR). Such a circuit rejects some of the pulse pile-up. Therefore, this circuit leads to counting losses. Determination of these counting losses is an important problem. In this work, a new method is suggested for the determination of the pulse pile-up reject.

RDL Kalıcı Bağlantı (PID)
https://rdlopen.rawdatalibrary.net/r/gumushane/15031Resmi atıflarda kullanılan kalıcı adres
DOI
https://doi.org/10.1063/1.4976408Yayıncı resmi DOI adresi
Tam Kayıt
AlanDeğerDil
dc.contributor.authorKarabidak, S.M.
dc.contributor.authorKaya, S.
dc.contributor.editorAkkus, B.
dc.contributor.editorDogan, G.S.
dc.contributor.editorOktem, Y.
dc.contributor.editorGuzelcimen, F.
dc.date.accessioned2026-10-22T10:23:29.194705597Z
dc.date.available2026-10-22T10:23:29.194705597Z
dc.date.issued2017
dc.description.abstractThe collection the charged particles produced by the incident radiation on a detector requires a time interval. If this time interval is not sufficiently short compared with the peaking time of the amplifier, a loss in the recovered signal amplitude occurs. Another major constraint on the throughput of modern x or gamma-ray spectrometers is the time required for the subsequent the pulse processing by the electronics. Two above-mentioned limitations are cause of counting losses resulting from the dead time and the pile-up. The pulse pile-up is a common problem in x and gamma ray radiation detection systems. The pulses pile-up in spectroscopic analysis can cause significant errors. Therefore, inhibition of these pulses is a vital step. A way to reduce errors due to the pulse pile-up is a pile-up inspection circuitry (PUR). Such a circuit rejects some of the pulse pile-up. Therefore, this circuit leads to counting losses. Determination of these counting losses is an important problem. In this work, a new method is suggested for the determination of the pulse pile-up reject.
dc.identifier.articlenumber060012
dc.identifier.doi10.1063/1.4976408
dc.identifier.isbn978-073541483-9
dc.identifier.issn0094-243X
dc.identifier.scopus2-s2.0-85015990102
dc.identifier.urihttps://rdlopen.rawdatalibrary.net/r/gumushane/15031
dc.identifier.volume1815
dc.identifier.wosWOS:000435205100064
dc.language.isoen
dc.publisherAmerican Institute of Physics Inc.
dc.relation.conferenceTurkish Physical Society 32nd International Physics Congress, TPS 2016
dc.relation.ispartofAIP Conference Proceedings
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.subject.classificationPhysics, Multidisciplinary
dc.subject.classificationPhysics
dc.subject.otherTIME
dc.titleThe determination of the pulse pile-up reject (PUR) counting for X and gamma ray spectrometry
dc.typeConference Object
dspace.entity.typePublication
rdl.citation.openalex1
rdl.citation.scopus1
rdl.citation.wos2
rdl.conference.date6 September 2016 through 9 September 2016
rdl.conference.placeKonacik-Bodrum
rdl.oa.statusclosed
rdl.source.indexSCOPUS
rdl.source.indexWOS
rdl.wos.indexConference Proceedings Citation Index - Science (CPCI-S)