Optimizing compton suppression systems: A parametric monte carlo study of NaI scintillator geometry

MakaleYalnız künye

Özet

Compton suppression systems are widely used to enhance gamma-ray spectral quality in applications such as nuclear safeguards and environmental radioactivity measurements. Although optimization of high-purity germanium (HPGe) detectors has been extensively reported, the geometric optimization of secondary sodium iodide [NaI(Tl)] scintillation detectors remains less studied. This work investigates the influence of annular and plug NaI(Tl) detector geometry—length, thickness, and radius—on Compton suppression performance using Monte Carlo simulations. Performance is evaluated in terms of the peak Compton suppression factor (CSFpeak) and total Compton suppression factor (CSFtotal). The results show that CSFpeak and CSFtotal increase with plug length (2–10.2 cm) and annular thickness (1–5.85 cm), with saturation observed beyond approximately 8–10 cm and 4–5 cm, respectively, due to reduced incremental photon absorption. The plug detector radius (3.72–5.1 cm) exhibits a nonlinear effect, with optimal performance achieved at 4.55–4.83 cm, beyond which gains are marginal (<5%). Suboptimal geometries (plug length <6 cm, annular thickness <3 cm, plug radius <4.28 cm) result in 20–60% lower suppression performance, increasing Compton continuum contributions in 662 keV spectra. The optimal configuration—plug length of 8–10 cm, plug radius of 4.55–4.83 cm, and annular thickness of 4–5 cm—achieves more than 90% of maximum CSFpeak and CSFtotal while maintaining a balance between performance and detector volume. These results provide quantitative design guidelines for optimizing NaI(Tl)-based Compton suppression systems for improved gamma-ray spectrometry performance.

Anahtar Kelimeler

RDL Kalıcı Bağlantı (PID)
https://rdlopen.rawdatalibrary.net/r/gumushane/14223Resmi atıflarda kullanılan kalıcı adres
Tam Kayıt
AlanDeğerDil
dc.contributor.authorKaya, Ali
dc.contributor.authorÇelik, Necati
dc.contributor.authorKaya, Selim
dc.date.accessioned2026-10-22T10:22:24.667447911Z
dc.date.available2026-10-22T10:22:24.667447911Z
dc.date.issued2026
dc.description.abstractCompton suppression systems are widely used to enhance gamma-ray spectral quality in applications such as nuclear safeguards and environmental radioactivity measurements. Although optimization of high-purity germanium (HPGe) detectors has been extensively reported, the geometric optimization of secondary sodium iodide [NaI(Tl)] scintillation detectors remains less studied. This work investigates the influence of annular and plug NaI(Tl) detector geometry—length, thickness, and radius—on Compton suppression performance using Monte Carlo simulations. Performance is evaluated in terms of the peak Compton suppression factor (CSFpeak) and total Compton suppression factor (CSFtotal). The results show that CSFpeak and CSFtotal increase with plug length (2–10.2 cm) and annular thickness (1–5.85 cm), with saturation observed beyond approximately 8–10 cm and 4–5 cm, respectively, due to reduced incremental photon absorption. The plug detector radius (3.72–5.1 cm) exhibits a nonlinear effect, with optimal performance achieved at 4.55–4.83 cm, beyond which gains are marginal (<5%). Suboptimal geometries (plug length <6 cm, annular thickness <3 cm, plug radius <4.28 cm) result in 20–60% lower suppression performance, increasing Compton continuum contributions in 662 keV spectra. The optimal configuration—plug length of 8–10 cm, plug radius of 4.55–4.83 cm, and annular thickness of 4–5 cm—achieves more than 90% of maximum CSFpeak and CSFtotal while maintaining a balance between performance and detector volume. These results provide quantitative design guidelines for optimizing NaI(Tl)-based Compton suppression systems for improved gamma-ray spectrometry performance.
dc.identifier.articlenumber114209
dc.identifier.doi10.1016/j.radphyschem.2026.114209
dc.identifier.eissn1879-0895
dc.identifier.issn0969-806X
dc.identifier.scopus2-s2.0-105043537542
dc.identifier.urihttps://rdlopen.rawdatalibrary.net/r/gumushane/14223
dc.identifier.volume249
dc.identifier.wosWOS:001819795200001
dc.language.isoen
dc.publisherElsevier Ltd
dc.relation.ispartofRadiation Physics and Chemistry
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.subjectCompton suppression system
dc.subjectGamma-ray spectrometry
dc.subjectHPGe
dc.subjectMonte Carlo simulation
dc.subjectNAI
dc.subject.classificationChemistry, Physical
dc.subject.classificationNuclear Science & Technology
dc.subject.classificationPhysics, Atomic, Molecular & Chemical
dc.subject.classificationChemistry
dc.subject.classificationPhysics
dc.subject.otherGamma ray spectrometers
dc.subject.otherGermanium compounds
dc.subject.otherIodine compounds
dc.subject.otherOptimal systems
dc.subject.otherPhosphorus compounds
dc.subject.otherRadioactivity
dc.subject.otherRadioactivity measurement
dc.subject.otherScintillation counters
dc.subject.otherSpectrometry
dc.subject.otherUltraviolet spectroscopy
dc.subject.othersodium iodide
dc.subject.otherCompton suppression
dc.subject.otherCompton suppression system
dc.subject.otherGamma rays spectrometries
dc.subject.otherGamma-rays
dc.subject.otherHigh purity germaniums
dc.subject.otherMonte Carlo's simulation
dc.subject.otherNAI
dc.subject.otherPerformance
dc.subject.otherSpectral quality
dc.subject.otherSuppression systems
dc.subject.otherArticle
dc.subject.othercompton effect
dc.subject.othergamma spectrometry
dc.subject.othergeometry
dc.subject.otherlight absorption
dc.subject.otherMonte Carlo method
dc.subject.otherscintillation
dc.subject.othersensitivity analysis
dc.subject.otherthickness
dc.subject.otheruncertainty
dc.subject.otherMonte Carlo methods
dc.titleOptimizing compton suppression systems: A parametric monte carlo study of NaI scintillator geometry
dc.typeArticle
dspace.entity.typePublication
rdl.citation.openalex0
rdl.citation.scopus0
rdl.citation.wos0
rdl.oa.statusclosed
rdl.source.indexSCOPUS
rdl.source.indexWOS
rdl.wos.indexScience Citation Index Expanded (SCI-EXPANDED)