A quantitative analysis on the effect of crystal parameters on full energy peak efficiency of a coaxial high purity Germanium detector for the energy range 50 keV-2.5 MeV

MakaleYayıncıda açık erişim

Özet

High Purity Germanium (HPGe) detectors are essential instruments in gamma-ray spectrometry, offering high sensitivity and exceptional energy resolution. The full-energy-peak (FEP) efficiency is a critical parameter that influences the accuracy of activity concentration measurements of radionuclides. This study examines the FEP efficiency of a coaxial HPGe detector, focusing on variations in crystal length, crystal radius, and crystal hole dimensions. For varying crystal lengths, the efficiency values show negligible differences at low energy (50 keV) but significant increases at higher energies, indicating that longer crystal lengths enhance efficiency. Similarly, the efficiency increases with larger crystal radii across all energy levels suggesting substantial efficiency gains even at low energies. However, variations in crystal hole radius and depth, exhibit minimal impact on FEP efficiency across all tested energy levels. These findings highlight that optimizing crystal length and radius is more crucial for changing detector efficiency compared to modifying hole dimensions, providing valuable insights for optimizing Monte Carlo models and detector design.

Anahtar Kelimeler

RDL Kalıcı Bağlantı (PID)
https://rdlopen.rawdatalibrary.net/r/gumushane/14161Resmi atıflarda kullanılan kalıcı adres
Tam Kayıt
AlanDeğerDil
dc.contributor.authorÇelik, Necati
dc.date.accessioned2026-10-22T10:22:19.808130245Z
dc.date.available2026-10-22T10:22:19.808130245Z
dc.date.issued2025
dc.description.abstractHigh Purity Germanium (HPGe) detectors are essential instruments in gamma-ray spectrometry, offering high sensitivity and exceptional energy resolution. The full-energy-peak (FEP) efficiency is a critical parameter that influences the accuracy of activity concentration measurements of radionuclides. This study examines the FEP efficiency of a coaxial HPGe detector, focusing on variations in crystal length, crystal radius, and crystal hole dimensions. For varying crystal lengths, the efficiency values show negligible differences at low energy (50 keV) but significant increases at higher energies, indicating that longer crystal lengths enhance efficiency. Similarly, the efficiency increases with larger crystal radii across all energy levels suggesting substantial efficiency gains even at low energies. However, variations in crystal hole radius and depth, exhibit minimal impact on FEP efficiency across all tested energy levels. These findings highlight that optimizing crystal length and radius is more crucial for changing detector efficiency compared to modifying hole dimensions, providing valuable insights for optimizing Monte Carlo models and detector design.
dc.identifier.articlenumber112171
dc.identifier.doi10.1016/j.radphyschem.2024.112171
dc.identifier.eissn1879-0895
dc.identifier.issn0969-806X
dc.identifier.scopus2-s2.0-85203062901
dc.identifier.urihttps://rdlopen.rawdatalibrary.net/r/gumushane/14161
dc.identifier.volume226
dc.identifier.wosWOS:001309884200001
dc.language.isoen
dc.publisherElsevier Ltd
dc.relation.ispartofRadiation Physics and Chemistry
dc.rightsinfo:eu-repo/semantics/closedAccess
dc.subjectFull energy peak efficiency
dc.subjectHPGe detectors
dc.subjectMonte Carlo simulation
dc.subject.classificationChemistry, Physical
dc.subject.classificationNuclear Science & Technology
dc.subject.classificationPhysics, Atomic, Molecular & Chemical
dc.subject.classificationChemistry
dc.subject.classificationPhysics
dc.subject.otherGermanium compounds
dc.subject.othergermanium
dc.subject.otherradioisotope
dc.subject.otherCrystal length
dc.subject.otherCrystal parameters
dc.subject.otherCrystal radius
dc.subject.otherEnergy
dc.subject.otherEnergy ranges
dc.subject.otherFull-energy peak efficiencies
dc.subject.otherGamma rays spectrometries
dc.subject.otherHigh-purity germanium detector
dc.subject.otherLower energies
dc.subject.otherMonte Carlo's simulation
dc.subject.otheraccuracy
dc.subject.otherArticle
dc.subject.otherdosimetry
dc.subject.othergamma spectrometry
dc.subject.otherMonte Carlo method
dc.subject.otherphoton
dc.subject.otherquantitative analysis
dc.subject.otherradiation protection
dc.subject.otherGamma ray spectrometers
dc.titleA quantitative analysis on the effect of crystal parameters on full energy peak efficiency of a coaxial high purity Germanium detector for the energy range 50 keV-2.5 MeV
dc.typeArticle
dspace.entity.typePublication
rdl.citation.openalex3
rdl.citation.scopus4
rdl.citation.wos5
rdl.oa.statusgreen
rdl.oa.statuspreprint
rdl.source.indexSCOPUS
rdl.source.indexWOS
rdl.wos.indexScience Citation Index Expanded (SCI-EXPANDED)