Determination of the dependence of HPGe virtual point detector location on source dimensions in 60 keV-2 MeV range using Monte Carlo simulation

MakaleAçık ErişimScopus SJRQ2

Özet

The virtual point detector (VPD) concept for an HPGe detector with 55% relative efficiency was confirmed for bulky sources having different diameters and heights. After reproducing experimental data, Monte Carlo simulation was used to investigate the dependence of VPD location on photon energy for various source diameters and heights in the energy range from 60 keV to 2 MeV. The simulations were confirmed using a soil reference source prepared by International Atomic Energy Agency (IAEA). It was seen that increasing photon energy makes significant changes in VPD location up to 600 keV. It was observed that VPD locations increase with increasing source diameter and height. It was also seen that VPD locations were approximately linearly dependent on source diameters and heights.

Anahtar Kelimeler

RDL Kalıcı Bağlantı (PID)
https://rdlopen.rawdatalibrary.net/r/gumushane/15609Resmi atıflarda kullanılan kalıcı adres
İndeks kayıtları
Kaydın indeks veri tabanlarındaki sayfası (aktarımda alınan kayıt numarasıyla). Erişim, indeksin kendi koşullarına bağlıdır.
Atıflar
Scopus
5
WoS
5
OpenAlex
6
Her kaynağın kendi saydığı toplam atıf; aktarım ve zenginleştirme sırasında alındı.
Tam Kayıt
AlanDeğerDil
dc.contributor.authorÇelik, N.
dc.date.accessioned2026-10-22T10:24:15.452995934Z
dc.date.available2026-10-22T10:24:15.452995934Z
dc.date.issued2012
dc.description.abstractThe virtual point detector (VPD) concept for an HPGe detector with 55% relative efficiency was confirmed for bulky sources having different diameters and heights. After reproducing experimental data, Monte Carlo simulation was used to investigate the dependence of VPD location on photon energy for various source diameters and heights in the energy range from 60 keV to 2 MeV. The simulations were confirmed using a soil reference source prepared by International Atomic Energy Agency (IAEA). It was seen that increasing photon energy makes significant changes in VPD location up to 600 keV. It was observed that VPD locations increase with increasing source diameter and height. It was also seen that VPD locations were approximately linearly dependent on source diameters and heights.
dc.identifier.articlenumberP08009
dc.identifier.doi10.1088/1748-0221/7/08/p08009
dc.identifier.issn1748-0221
dc.identifier.issue8
dc.identifier.scopus2-s2.0-84865210243
dc.identifier.urihttps://rdlopen.rawdatalibrary.net/r/gumushane/15609
dc.identifier.volume7
dc.identifier.wosWOS:000308869800020
dc.language.isoen
dc.publisherIOP PUBLISHING LTD
dc.relation.ispartofJournal of Instrumentation
dc.rightsinfo:eu-repo/semantics/openAccess
dc.subjectDetector modelling and simulations I (interaction of radiation with matter interaction of photons with matter interaction of hadrons with matter etc)
dc.subjectGamma detectors (scintillators CZT HPG HgI etc)
dc.subjectModels and simulations
dc.subject.classificationInstruments & Instrumentation
dc.titleDetermination of the dependence of HPGe virtual point detector location on source dimensions in 60 keV-2 MeV range using Monte Carlo simulation
dc.typeArticle
rdl.citation.openalex6
rdl.citation.scopus5
rdl.citation.wos5
rdl.oa.statusbronze
rdl.oa.urlhttps://iopscience.iop.org/article/10.1088/1748-0221/7/08/P08009/pdf
rdl.scopus.quartileQ2
rdl.source.indexSCOPUS
rdl.source.indexWOS
rdl.wos.indexScience Citation Index Expanded (SCI-EXPANDED)